X-Cite Optical Power Measurement System
The X-Cite®Optical Power Measurement System is designed for measuring optical power in watts at the specimen level that enables consistent and repeatable illumination throughout research experiments, and also assist in equipment set-up and troubleshooting. The system includes the X-Cite XP750 sensor which has a low, sleek profile that fits on the microscope stage and offers the versatility of measuring power output from an X-Cite microscope illuminator or any other epi-fluorescence light source. By fitting a wide dynamic range into a compact sensor, we have made it easier than ever to obtain this critical information.
Broad Dynamic Range
Calibrated for use at any wavelength between 320 nm and 750 nm, X-Cite XP750 is compatible with a full range of filters. With sensitivity to power levels from 5 µW to 500 mW, it is appropriate for applications using both low and high intensity illumination. This makes it suitable for use on standard, confocal, DSU and other microscope configurations.
Versatile and Convenient
X-Cite XP750 measures light right on the microscope stage so it can be used with any epi-fluorescence light source including: HBO / mercury, metal halide or xenon lamps, lasers and LEDs. With hundreds of wavelength choices, X-Cite XR2100 allows you to define ‘favorite wavelengths’ to correspond to your most frequently used sources and filters.
Repeatable Results
By measuring and recording power output in absolute units (watts) with X-Cite, illumination levels used in an experiment can always be repeated, no matter how light sources, light guides, filters and other optical components change over time. This unique capability is critical for reducing post-experiment image processing time; making accurate, quantitative image comparisons; and compiling complete, experiment documentation.
TECHNICAL SPECIFICATIONS |
X-Cite XR2100 |
X-Cite XP750 |
Includes |
Handheld power meter, adapter for 3mm light guide, software CD, cables, user manual |
Objective plane power sensor with cable / connector for X-Cite XR2100 |
Power Range |
50 mW-10 W |
5 µW-500 mW |
Measurement Resolution |
0.1 mw-0.01 w |
0.01 µW-1 mW |
Uncertainty*** |
±5% |
±6% |
Response Time |
1 s |
600 ms (initial), 3 s (to ensure stable reading) |
Calibration |
可追溯到nist * |
Traceable to NRC** |
Wavelength Range |
340 nm-675 nm |
320 nm-750 nm |
灯类型/ Light Source Compatibility |
X-Cite 200DC, X-Cite exacte, X-Cite 120 Series (using 3 mm light guide input port) |
X-Cite 200DC, X-Cite exacte, X-Cite 120 Series, Mercury / HBO, Metal Halide, Xenon, LED, Laser |
Objective Compatibility |
Not applicable |
4X-63X; air coupled, with FOV diameters less than 10mm |
Display |
3位数LCD,背光 |
通过X-CITE XR2100 |
Wavelength Selection |
Not Applicable |
1nm increments using up / down buttons on X-Cite XR2100 or PC interface |
Data Capacity |
在手持设备上存储100多个读数,或直接记录到PC接口中;导出电子表格兼容格式 |
通过X-CITE XR2100 |
PC Controls |
View / change settings, download / export stored data |
查看/更改设置,定义喜爱的波长,多个目标/过滤器/强度设置的记录数据,下载/导出存储数据 |
命令协议 |
RS232 via USB virtual COM port |
通过X-CITE XR2100 |
电源供应 |
2 x 3.6 V Lithium Battery |
通过X-CITE XR2100 |
Weight |
1 lb (450 g) |
2.9 oz (82 g) |
Dimensions (without cover) |
7.5“x 4.5”x 2“(19 cm x 11.5 cm x 5 cm) |
3" x 1" x 0.35" (75 mm x 2 5mm x 9 mm) |
Worldwide Certifications |
CE marked, RoHS compliant |
通过X-CITE XR2100 |
Warranty |
1 year |
1 year |
专利 |
X-Cite Optical Power Measurement System incorporates technology protected by the following patents: US#6,437,861; US#7,335,901 |
*NIST – National Institute of Standards and Technology
** NRC - 国家研究理事会
***每12个月建议X-CITE XR2100和X-CITE XP750的校准。联系我们获取更多信息。
TECHNICAL SPECIFICATIONS |
X-Cite XR2100 |
X-Cite XP750 |
Includes |
Handheld power meter, adapter for 3mm light guide, software CD, cables, user manual |
Objective plane power sensor with cable / connector for X-Cite XR2100 |
Power Range |
50 mW-10 W |
5 µW-500 mW |
Measurement Resolution |
0.1 mw-0.01 w |
0.01 µW-1 mW |
Uncertainty*** |
±5% |
±6% |
Response Time |
1 s |
600 ms (initial), 3 s (to ensure stable reading) |
Calibration |
可追溯到nist * |
Traceable to NRC** |
Wavelength Range |
340 nm-675 nm |
320 nm-750 nm |
灯类型/ Light Source Compatibility |
X-Cite 200DC, X-Cite exacte, X-Cite 120 Series (using 3 mm light guide input port) |
X-Cite 200DC, X-Cite exacte, X-Cite 120 Series, Mercury / HBO, Metal Halide, Xenon, LED, Laser |
Objective Compatibility |
Not applicable |
4X-63X; air coupled, with FOV diameters less than 10mm |
Display |
3位数LCD,背光 |
通过X-CITE XR2100 |
Wavelength Selection |
Not Applicable |
1nm increments using up / down buttons on X-Cite XR2100 or PC interface |
Data Capacity |
在手持设备上存储100多个读数,或直接记录到PC接口中;导出电子表格兼容格式 |
通过X-CITE XR2100 |
PC Controls |
View / change settings, download / export stored data |
查看/更改设置,定义喜爱的波长,多个目标/过滤器/强度设置的记录数据,下载/导出存储数据 |
命令协议 |
RS232 via USB virtual COM port |
通过X-CITE XR2100 |
电源供应 |
2 x 3.6 V Lithium Battery |
通过X-CITE XR2100 |
Weight |
1 lb (450 g) |
2.9 oz (82 g) |
Dimensions (without cover) |
7.5“x 4.5”x 2“(19 cm x 11.5 cm x 5 cm) |
3" x 1" x 0.35" (75 mm x 2 5mm x 9 mm) |
Worldwide Certifications |
CE marked, RoHS compliant |
通过X-CITE XR2100 |
Warranty |
1 year |
1 year |
专利 |
X-Cite Optical Power Measurement System incorporates technology protected by the following patents: US#6,437,861; US#7,335,901 |
*NIST – National Institute of Standards and Technology
** NRC - 国家研究理事会
***每12个月建议X-CITE XR2100和X-CITE XP750的校准。联系我们获取更多信息。